The spectra of light treatment equipment that are used in treatment of winter SAD and hyperbilirubinaemia were measured using a spectrometer. There were five bright light instruments that are used for winter SAD treatment and four blue light instruments that are used for treatment of hyperbilirubinaemia, which were measured. The measurements were carried out in circumstances equivalent to the normal use. The spectra were measured in wavelength range of 200...800 nm. Two instruments were additionally measured in wawelength range of 250...400 nm in order to compare the results.
The results were investigated using the limit values for exposure to ultraviolet radiation and visible light given by ACGIH, IRPA, CIE and Ministry of Health and Social Affairs. The measured instruments were all proved to be safe to use. The maximum permissible exposure durations were between 25...1025 hours for the bright light instruments and between 4...7716 hours for the blue light instruments, when UV-radiation was considered. In the range of blue light the permissible exposure durations with no eye protection were between 13...111 hours for the bright light instruments and between 6...38 hours for the blue light instruments-------
7. RESULTS
7.1 Computational values
Tabel 7 shows the maximum exposure time for UV-radiation in interval 200 400 nm according to ACGIH:s recommendations. Tabell 8 innehåller motsvarande värden beräknade enligt CIE:s rekommendationer. Värdena i tabell 9 bygger på IRPA:s direktiv gällande exponering för UV-strålning. Värdena har beräknats för strålning i intervallet 200 400 nm. Tabell 10 visar radiansvärdena i intervallet 400 700 nm som fåtts genom att mäta irradiansvärdena. Radianvärdena har viktats med funktionen av ljusets farlighet. Utgående från detta har tillåtna exponeringstider beräknats.
Tabell 7. Uppmätta irradianser i intervallet 200 400 nm och motsvarande maximala exponeringstider enligt ACGIH:s rekommendationer. Ee = totalirradians, Eeff = effektiv irradians. Exponeringstiderna har beräknats med hälp av den effektiva radiansen. De värden som beräknats med teflondiffusor utgående från mätningar med Drägers lampa och den taklampa som är jämförbar med denna finns på den undre raden för respektive ljuskälla.
Tabell 7 (Valotaina Ltd has made the cursives and strenghnesses to get the table clearer)
Ee (W/cm2) Eeff (W/cm2) tmax(seconds) tmax(hours) HYKS 6,93E-7 2,41E-10 12 448 133 3458 Medela 3,58E-6 3,22E-10 9 316 770 2588 Dräger 4,68E-5
6,35E-52,64E-9
3,4E-91 136 364
882 353316
245Wallaby 8,85E-7 1,08E-10 27 777 778 7716 Aleksi (Philips) 6,00E-6 2,76E-9 1 086 957 302 True-Lite 6,13E-6 1,28E-8 234 375 65 Ceiling lamp (Ceiling-Sunshine, designmodel)
5,26E-6
1,13E-58,27E-9
1,79E-8362 756
167 597101
47Desktop lamp (Desktop-Sunshine)
7,13E-6 8,13E-10 3 690 037 1025 Small (Tabletop-Sunshine)
7,60E-6 9,37E-10 3 201 708 889 Tabell 7 visar att totalirradianserna ligger märkbart under ACGHI:s säkerhetsvärden. Enligt ACGIH får totalirradiansen inte överstiga 1,0 mW/cm2 om ögat är oskyddat och strålningen pågår i mer än 103 s eller ca 16 minuter. De uppmätta totalirradianserna i de apparater som undersöktes var av storleksordningen 10 µW/cm2. ACGIH:s tillåtna gränsvärden för effektiva irradianser beror på exponeringstiden. Den längsta exponeringstiden som getts ett gränsvärde är 8 timmar med en tillåten UV-exponering på 0,1 µW/cm2. Tabell 8 visar att även detta värde överskrids av alla ljusterapilampor, och de närmaste värdena är av storleksordningen 0,01 µW/cm2. De tillåtna exponeringstiderna är långa på grund av de låga irradianserna; från tiotals till tusentals timmar.